Why do arterial and PV traces look different on ECMO or Impella?

Continuous flow can sustain mean pressure with reduced native pulsatility.

Overview

ECMO and Impella add continuous flow, while native ventricular ejection supplies the pulsatile part. Strong support with little native ejection can therefore produce a nearly flat arterial trace around a supported mean pressure.

Impella drains the LV toward the aorta; peripheral V-A ECMO bypasses the lungs and can increase LV loading. The modeled chamber pressures and PV loops respond differently to those configurations.

During ECMO-supported VF or a zero-rate state, vasoactive drugs and sedative vascular effects still change SVR and supported mean pressure. Native ejection remains absent: a pressure rise is not rhythm conversion or restoration of a pulse.

The model is not meant to force every supported tracing to a completely flat line. Residual native ejection, rhythm, loading and support level determine how much pulse remains.